ORGANIC SOLVENT PURIFICATION SYSTEM
Technical Specification
We provides customers with a fast, safe and easy-to-operate organic solvent purification system. This product offers laboratories a system that meets the highest safety-level requirements and can be customized according to specific needs.
The organic solvent purification system delivers solvent by pressurizing the liquid surface in the reservoir tank with a working gas (99.99% nitrogen or higher-purity gas is recommended). The solvent flows through the pipeline system into dual purification columns that remove water and oxygen. The ultra-pure solvent then flows into an evacuated solvent receiving bottle, a reactor, or a glovebox.
After the spent purification material has been processed by the customer, We can provide filling and activation services with new packing material.
The system is a closed-circuit operation system and mainly consists of the following components:
• Working gas (N₂ or Ar) supply
• Solvent reservoir tank (stainless steel, 1.4301/US 304)
• Dual purification columns (water and oxygen removal)
• Vacuum pump (oil-free diaphragm pump)
• Organic solvent collection port
• Exhaust / solvent vapor removal port
• Option of direct collection, collection inside a glovebox, or both
• Pressure reducing valve and flow control valve
• Vacuum/working gas filling valve, and vacuum/working gas or solvent control valve
• Receiving bottle support rod
• Safety fireproof cabinet (optional)
• Casters and leveling feet

|
Solvent Category |
Solvents |
|
Aromatic and aliphatic hydrocarbons |
Pentane, n-hexane, cyclohexane, n-heptane, toluene, benzene |
|
Ethers |
Diethyl ether, tetrahydrofuran, dimethyl ether |
|
Chlorinated solvents |
Dichloromethane, chloroform, chlorobenzene |
|
Amine solvents |
Triethylamine, pyridine, N,N-diisopropylethylamine |
|
Alcohols |
Methanol, ethanol |
|
Other common solvents |
Acetonitrile, dimethylformamide, dimethyl sulfoxide, acetone |
Note: This solvent list is not exhaustive; other similar compounds can be provided upon request.
|
Parameter |
SolvPurer A3/G3 |
SolvPurer A5/G5 |
SolvPurer A7/G7 |
|
Model |
SolvPurer A3/G3 |
SolvPurer A5/G5 |
SolvPurer A7/G7 |
|
Number of Solvents |
3 |
5 |
7 |
|
System Control |
Manual (upgradable to PLC automatic control system) |
Manual (upgradable to PLC automatic control system) |
Manual (upgradable to PLC automatic control system) |
|
Control |
Manual collection of purified solvent |
Manual collection of purified solvent |
Manual collection of purified solvent |
|
Material |
Stainless steel (1.4301/US 304) |
Stainless steel (1.4301/US 304) |
Stainless steel (1.4301/US 304) |
|
Leveling Feet |
Yes |
Yes |
Yes |
|
System Assembly |
Stainless steel tubing with five-way manual valves |
Stainless steel tubing with five-way manual valves |
Stainless steel tubing with five-way manual valves |
|
Work Table |
Organic solvent receiving support plate with height adjustment and fixing mechanism |
Organic solvent receiving support plate with height adjustment and fixing mechanism |
Organic solvent receiving support plate with height adjustment and fixing mechanism |
|
Casters |
Yes |
Yes |
Yes |
|
Vacuum Pump |
Oil-free diaphragm pump, pumping speed 20 L/min; vacuum less than 8 mbar |
Oil-free diaphragm pump, pumping speed 20 L/min; vacuum less than 8 mbar |
Oil-free diaphragm pump, pumping speed 20 L/min; vacuum less than 8 mbar |
|
Moisture Meter (optional) |
Titration method: coulometric Karl Fischer; detection range: 10 μg – 99 mg (1 ppm – 100%) |
Titration method: coulometric Karl Fischer; detection range: 10 μg – 99 mg (1 ppm – 100%) |
Titration method: coulometric Karl Fischer; detection range: 10 μg – 99 mg (1 ppm – 100%) |
|
Solvent Flow Rate |
< 1 L/min |
< 1 L/min |
< 1 L/min |
|
Power Supply (vacuum pump and PLC controller) |
230 V/50–60 Hz, 10 A; 115 V/50–60 Hz, 15 A |
230 V/50–60 Hz, 10 A; 115 V/50–60 Hz, 15 A |
230 V/50–60 Hz, 10 A; 115 V/50–60 Hz, 15 A |
|
Fireproof Safety Cabinet |
Optional |
Optional |
Optional |
|
Solvent Vapor Removal / Exhaust Port |
With relevant interfaces |
With relevant interfaces |
With relevant interfaces |
|
Operation |
Closed-circuit operation system |
Closed-circuit operation system |
Closed-circuit operation system |
|
Solvent Collection (receiving purified solvent) |
Direct collection, collection inside a glovebox, or both |
Direct collection, collection inside a glovebox, or both |
Direct collection, collection inside a glovebox, or both |
|
Working Gas |
Nitrogen or argon (gas purity of 99.99% or higher recommended) |
Nitrogen or argon (gas purity of 99.99% or higher recommended) |
Nitrogen or argon (gas purity of 99.99% or higher recommended) |
|
Solvent Treatment System |
Each solvent has its own independent system |
Each solvent has its own independent system |
Each solvent has its own independent system |
|
Solvent Line Material |
Stainless steel tubing (1.4301/US 304) |
Stainless steel tubing (1.4301/US 304) |
Stainless steel tubing (1.4301/US 304) |
|
Inlet Pressure |
Set pressure between 4.0 bar and 6.0 bar |
Set pressure between 4.0 bar and 6.0 bar |
Set pressure between 4.0 bar and 6.0 bar |
|
Working Pressure |
Set pressure between 0.3 bar and 0.5 bar |
Set pressure between 0.3 bar and 0.5 bar |
Set pressure between 0.3 bar and 0.5 bar |
|
Purification Column Type |
Dual-column type |
Dual-column type |
Dual-column type |
|
Purification Column Structure |
Two purification columns per solvent (1.4301/US 304) |
Two purification columns per solvent (1.4301/US 304) |
Two purification columns per solvent (1.4301/US 304) |
|
Purification Column Size |
Φ100 mm × 700 mm |
Φ100 mm × 700 mm |
Φ100 mm × 700 mm |
|
Particle Filter |
Columns equipped with pre-treatment filter |
Columns equipped with pre-treatment filter |
Columns equipped with pre-treatment filter |
|
Packing Material Type |
Depends on solvent |
Depends on solvent |
Depends on solvent |
|
Solvent Capacity Processed per Column Set |
≥ 800 L |
≥ 800 L |
≥ 800 L |
|
Column Activation Method |
Pre-activated before delivery |
Pre-activated before delivery |
Pre-activated before delivery |
|
Achievable Solvent Purity |
Depends on the solvent type |
Depends on the solvent type |
Depends on the solvent type |
|
Solvent Reservoir Tank — Material |
Stainless steel reservoir tank (1.4301/US 304) |
Stainless steel reservoir tank (1.4301/US 304) |
Stainless steel reservoir tank (1.4301/US 304) |
|
Solvent Reservoir Tank — Capacity |
20 L |
20 L |
20 L |
|
Solvent Reservoir Tank — Features |
Double-valve shut-off device, quick couplings, special line safety valve and filling port |
Double-valve shut-off device, quick couplings, special line safety valve and filling port |
Double-valve shut-off device, quick couplings, special line safety valve and filling port |
|
Solvent Reservoir Tank — Tubing |
Stainless steel tubing with ferrule-type fittings |
Stainless steel tubing with ferrule-type fittings |
Stainless steel tubing with ferrule-type fittings |